TOC Analyzer

TOC-50 TOC Multi-point Analyzer, GMP

TOC-50

The TOC-50 Total Organic Carbon Analyzer consists of a TOC-50 Total Organic Carbon (TOC) data terminal and many total organic carbon (TOC) detection unit.

More details

More info

Product Description:

TOC-type Total Organic Carbon Analyzer is Labgeni independent research and developed instrument forthe determinationtotal organic carbon concentrationin the water samples. It can detect the TOC concentration from 1ppb to 1000ppb of the water samples, with high sensitivity and accuracy .

 

TOC-50 Total Organic Carbon Analyzer consists of a TOC-50 Total Organic Carbon (TOC) data terminal and many total organic carbon (TOC) detection unit. The data terminal is responsible for collecting and displaying the data of each detection unit, with data storage, printing, alarm and other functions. Each test unit has its own oxidation detector, respectively detection the TOC of various water samples monitoring point.

 

Application areas:

Can be used to detect the concentration of total organic carbon in purified water, water for injection and deionized water in the pharmaceutical industry. It can monitor the water system of the pharmaceutical industry, the ultrapure water preparation system and the wafer process of the semiconductor industry, the preparation process of deionized water in power plant, can also be used for semiconductor industry, power plants, scientific research units, laboratories and other ultra-pure water TOC detection.

 

Main feature:

1, Adopt super large 10-inch touch screen, can control maximum 8 sets detection unit;

2, Adopt 485 communication interface;

3, Query according to the time and delete the history;

4, Detection according to the UV static oxidation principle;

5, Inner diameter of the inlet and outlet water tube is 1mm Teflon tube, configure  50 pcs quick connector;

6, Ultra-low detection limit;

7, Best data stability.

8, Matching micro printer, real-time printing.

 

Technical Parameters:

Model

TOC-50

TOC detection range

0.001mg/L-1.5mg/L(1500ppb)

Conductivity detection range

0.055 µS/cm ~6.000 µS/cm

TOC detection limit

1 µg/L(0.01mg/L)

TOC detection accuracy

±5%

Response time

Within 30 min

Repeatability error

≤ 3%

Zero point drift

±5%

Measuring Range drift

±5%

Sample temperature

0~100℃

Ambient temperature

5~65℃

Relative humidity

≤ 85%

Power supply

220V±10%

Power frequency

50Hz±1Hz

Detection unit rated power

50W

9 other products you may be interested: