Performance Parameter:
|
Power requirements / power |
170-240V AC, 50/60Hz;100W |
170-240V AC, 50/60Hz;85w |
170-240V AC, 50/60Hz;100W |
|
Maximum operating pressure |
8000mbar |
8000 mbar |
8000mbar |
|
Minimum inlet pressure |
100mbar |
100 mbar |
100mbar |
|
Dimension |
400(W) x 380(D) x 100(H1) x 235(H2) |
240(W) x 380(D1) x 280(D2) x 220(H1) |
400(W) x 380(D) x 100(H1) x 235(H2) |
|
Measuring Range |
500-6900mbar |
500-6900mbar |
500-6900mbar |
|
Measuring Accuracy |
Net Volume Test:± 4%;Bubble Point:± 50mbar ;Diffusion Current:± 4% ; |
Sensitivity:± 1.0 mbar ; Bubble Point:± 75 mbar |
Net Volume Test:± 4%;Bubble Point:± 50mbar ;Diffusion Current:± 4% ; |
|
Operating condition |
Ambient Temperature:+5℃ ~ +35℃;Relative Humidity:10-80% |
Ambient Temperature:+5℃ ~ +35℃;Relative Humidity:10-80% |
Ambient Temperature:+5℃ ~ +35℃;Relative Humidity:10-80% |
|
Test time consuming |
Net Volume Test:5 min±2min; Diffusion Current Test:10 min±2min; |
Bubble point test :15 min±2min; bubble point scanning : 20min±2min;Diffusion Current:10min±2min |
Net Volume Test:5 min±2min; Diffusion Current Test:10 min±2min; |
|
Printing function |
W line Chinese printing, Output test parameters and test results |
Chinese printing, Output test parameters and test results |
W line Chinese printing, Output test parameters and test results |
|
History recording function |
500 group Test result( Including test curves ) |
Store 20 groups of test results |
500 group Test result(Including test curves ) |
|
Display |
size:5.7″ TFT; homochromy |
size:73*39mm; homochromy |
Size:5.7″ TFT; homochromy |
|
serial connection |
serial port :RS232 |
|
serial port :RS232 |
|
Language |
Chinese |
Chinese |
Chinese |
Main Characteristic:
Online testing ,does not interfere the downstream aseptic conditions
It has system self-check function, with an error report function for the system tightness and wrong operation;
It is microcomputer control of automatic detection, with many kinds of test functions such as bubble point scanning, fast bubble point test and pressure holding method;
Real-time print of test results;
It can be used to test the single core filter and multi-core filter.