PT-D445-3007 Asphalt Kinematic Viscosity Apparatus measures absolute viscosity of viscosity graded paving bitumen (IS: 73: 2006) at 60 C in accordance with IS: 1206 (Part II) (similar to ASTM D 2171), which uses a vacuum capillary viscometer. It should be noted that any other viscometer such as rotational viscometer 
Cannot be used to measure the absolute viscosity at 60 C since the bitumen in nonNewtonian in nature at this temperature. While obtaining quotations specify the following requirements. Although Asphalt Institute vacuum viscometer and Modified Koppers vacuum viscometer can also be used, the following refers to Cannon-Manning vacuum viscometer only. 

Technical parameters 

  1. Power supply: AC 220V± 10%, 50 Hz; 
  2. Heating power: Two grades; 1000 W for auxiliary heating and 600 W for temperature controlling; 
  3. Bath temperature: Ambient temperature~180.0° C; 
  4. Temperature controlling accuracy: ± 0.1° C; 
  5. Mercury thermometer: 
  6. Rod type 
  7. Scale division: 0.1 ° C; 
  8. Measurement range: 100~150° C or 150~200° C; 
  9. Bath cubage: Not less than 23 L; 
  10. Sample quantity: 3 piece at the same time; 
  11. Stirring motor: 6 W, 1200 RPM; 
  12. Ambient temperature: -10° C~+35° C; 
  13. Relative humidity: <85%; 
  14. Temperature sensor: Industrial platinum resistance, Pt100; 
  15. Total Power consumption: Not more than 1800 W. 
  16. Capillary viscometer: 7 pieces of Cannon-Fenske reserve-flow capillary viscometer tubes in a group; Their models are 200, 300, 350, 400, 450, 500, 600 respectively (Corresponding diameters are 1.02, 1.26, 1.48, 1.88, 2.20, 3.10 and 4.00 mm). The users should prepare Ubbelohde type viscometer tubes by themselves.