X-ray Fluorescence

X-Fluorescence Sulfur Calcium and Ferrum Analyzer

EA-XF3300

EA-XF3300 X-fluorescence sulfur calcium and ferrum analyzer could be applied in building material, thermoelectricity, mine, plaster, coal, etc, can quickly analyze the percentage content of SO3 ,CaO and Fe2O3  in the material .

More details

More info

1. Main Usage
EA-XF3300 X- fluorescence sulfur calcium and ferrum analyzer could be applied in building material, thermoelectricity, mine, plaster, coal, etc, can quickly analyze the percentage content of SO3 ,CaO and Fe2O3  in the material . Due to its fast analysis, it is already widely used in cement industry. It could monitor real time the change condition of these three Ingredients in production process, so as to adjust raw material ratio in time, laying a solid foundation for producing qualified raw material, clinker, cement. It is used in
thermal power plant, which could detect the calcium content in limestone used in
desulfurization, identify the quality of factory raw material; Detect the sulfur content
in desulfurization ash, identify desulfurization effects; Could detect total sulfur content
of factory coal.

2. Main Feature
1) Analysis range reaches to 15% width,  which breaks the only 5%-7% measuring range of same type instrument in many year.
2) Short analysis time, percentage content of SO3 , CaO is given in 30S,
estimate the mixing amount of in the cement.
3) Large data storage capacity, content result, instrument self-checking data could be queried, avoid printing supplier.
4) Micro computer integrated as a whole, compact structure, beautiful appearance
5) Don’t destroy samples when instrument detects, sample can be used  repeatedly .
6) Without any chemical reagent, no "three wastes" emissions, don’t contain radioactive source, low-power, comply with the requirement of environmental protection and energy saving.
7) Large screen LCD display, entirely in English menu prompt operation, convenient to use.
8) Instrument execute GB/T19140 《the general principles of cement X ray fluorescence analysis》

3.   Technical Indicator
1) Analysis range: 0.01%~100%

2) Analysis precision : Standard deviation SSO3≤0.04%、 SCaO≤0.05%,SFe2O3≤0.04%

3) Analysis width: CaO: min~max≤15% , for example CaO :40%~55%  in the cement, selected according to calibration working curve.
4) Analysis time: n* 30S( n is 1-5 natural number)
5) Using condition: power supply AC200V-240V;environment temperature 0-40℃;Relative humidity<85%
6) Overall power consumption: <30W
7) Overall dimension: 468*368*136mm
8) Total weight: 13.8kg

 

 

7 other products you may be interested: